MSR145 Data Logger with Four Analogue Inputs and 20 cm Cable
Compact data logger with four analogue voltage inputs, each configured for 0–3 V, on a 20 cm cable. Records at up to 50 readings per second and stores more than two million measured values.
This MSR145 is configured to record four analogue voltage signals from 0 to 3 V through the E3333/020 connection. The 20 cm cable connects the data logger to compatible external signal sources or sensors; those external sensors are not included. Recording intervals can be selected from 50 readings per second to one reading every 12 hours. More than two million measured values can be stored internally.
The B15 polycarbonate housing contains a 230 mAh rechargeable battery. IP60 means the housing is dust-tight, with no water-protection rating. USB provides charging and data transfer, and the included MSR PC Software Pro supports configuration and evaluation of the recorded signals. This configuration is supplied with four 0–3 V inputs; it does not include the other voltage or current ranges offered elsewhere in the MSR145 range.
Benefits
- Four analogue channels configured for 0–3 V signals
- 20 cm connection cable
- Recording rate up to 50 Hz
- Compact, battery-powered acquisition with USB recharge
- Included MSR PC Software Pro for setup and evaluation
CONFIGURE YOUR OWN DEVICE >>
Technical specifications
| Manufacturer | MSR Electronics GmbH |
|---|---|
| Product code | 145B15 & E3333/020 |
| Housing | Polycarbonate, not encapsulated |
| Protection rating | IP60 |
| Battery | 230 mAh rechargeable lithium-polymer |
| Charging | USB |
| Dimensions of data logger (W × H × L) | 27 × 16 × 53 mm |
| Approximate data logger housing weight | 20 g; connected sensor assemblies add weight |
| Operating temperature of data logger | −20 to +65 °C |
| Storage conditions | +5 to +45 °C; 10–95 % RH, non-condensing |
| Interfaces | USB |
| Internal memory | Over 2,000,000 measured values |
| PC requirement | Windows 10 or higher for current MSR PC Software Pro |
| Analogue inputs | 4 × 0–3.0 V |
| Analogue resolution | 12 bit |
| Analogue interval | 0.02 second to 12 hours |
| External component E3333/020 | 1 × E3333/020; cable 20 cm |
| Humidity accuracy H3/FH3 | ±2 % RH (10–80 % RH, +25 °C); ±3 % RH (80–100 % RH, +25 °C) |
| Humidity sensor temperature limits | External: −40 to +125 °C; internal: −20 to +65 °C |
| Humidity interval | 1 second to 12 hours |
| Humidity sensor additional channel | Integrated temperature |
| Absolute air pressure range | 10–2,000 mbar absolute |
| Air pressure accuracy | ±2 mbar at 750–1,100 mbar absolute and +25 °C |
| Pressure sensor additional channel | Integrated temperature |
| Air pressure interval | 1 second to 12 hours |
| Light range | 0–83,000 lx |
| Light sensitivity | Maximum at 550 nm |
| Light interval | 1 second to 12 hours |
Included configuration
- MSR145 data logger in the specified housing with rechargeable battery
- Selected sensors and connections listed in this configuration
- MSR PC software (download)
- USB connection cable
Requirements and accessories
- Windows 10 or higher with USB connection for current MSR PC Software Pro
- Suitable external signal sources; observe the configured input voltage range