MSR145 Strain-Gauge Data Logger with 20 cm Connection

SKU: 145B7 & FD/020
Regular price 510.00 € (ex VAT)

Data logger with one 20 cm connection for a compatible full-bridge strain gauge, a 3 V bridge supply and a 1000 mAh rechargeable battery. Stores more than two million measured values; the strain gauge is not included.

 

This MSR145 configuration is equipped with one FD/020 connection for a full-bridge strain gauge on a 20 cm cable. The connection supplies a 3 V bridge voltage and records at intervals from one second to 12 hours. It provides the recording instrument and connection for a compatible strain gauge; the gauge itself is not part of this configuration.

The B7 housing contains a 1000 mAh rechargeable battery and memory for more than two million measured values. Its polycarbonate enclosure has an IP60 rating, meaning it is dust-tight but has no water-protection rating. Charge and read out the data logger through USB, and use the included MSR PC Software Pro for setup and measurement analysis. Compatibility of the intended bridge must be checked against the confirmed FD specification before use.

Benefits

  • One dedicated connection for a full-bridge strain gauge
  • 20 cm cable and 3 V bridge supply
  • 1000 mAh rechargeable battery
  • More than two million measured values in internal memory
  • USB configuration and analysis with MSR PC Software Pro



CONFIGURE YOUR OWN DEVICE >>

 

Technical specifications

Manufacturer MSR Electronics GmbH
Product code 145B7 & FD/020
Housing Polycarbonate, not encapsulated
Protection rating IP60
Battery 1000 mAh rechargeable lithium-polymer
Charging USB
Dimensions of data logger (W × H × L) 39 × 23 × 72 mm
Approximate data logger housing weight 56 g; connected sensor assemblies add weight
Operating temperature of data logger −20 to +65 °C
Storage conditions +5 to +45 °C; 10–95 % RH, non-condensing
Interfaces USB
Internal memory Over 2,000,000 measured values
PC requirement Windows 10 or higher for current MSR PC Software Pro
Strain gauge connection Full bridge; rated output 1–4 mV/V; bridge voltage 3 V
Strain gauge interval 1 second to 12 hours
External component FD/020 1 × FD/020; cable 20 cm
Relative humidity range 0–100 % RH
Humidity accuracy H3/FH3 ±2 % RH (10–80 % RH, +25 °C); ±3 % RH (80–100 % RH, +25 °C)
Humidity sensor temperature limits External: −40 to +125 °C; internal: −20 to +65 °C
Humidity interval 1 second to 12 hours
Humidity sensor additional channel Integrated temperature
Absolute air pressure range 10–2,000 mbar absolute
Air pressure accuracy ±2 mbar at 750–1,100 mbar absolute and +25 °C
Pressure sensor additional channel Integrated temperature
Air pressure interval 1 second to 12 hours
Acceleration 3 axes; ±15 g
Acceleration accuracy ±0.15 g (0–5 g), ±0.25 g (5–10 g), ±0.45 g (10–15 g), at +25 °C
Acceleration recording interval 0.02 second to 12 hours

 

Included configuration

  • MSR145 data logger in the specified housing with rechargeable battery
  • Selected sensors and connections listed in this configuration
  • MSR PC software (download)
  • USB connection cable

Requirements and accessories

  • Windows 10 or higher with USB connection for current MSR PC Software Pro
  • Compatible full-bridge strain gauge; the product code specifies the connection, not a strain gauge sensor

Documents and resources